Medgraphics Clinical Research
 
  Clinical Research Home | About Us | Products | Contact Us | Corporate

The Ultima CCM provides real solutions for:

  • Bias Flow
  • Pressure Support
  • Fluctuating FiO2
  • ICU Bedside Access

The revolutionary Ultima CCM provides a simple and easy patient interface to make metabolic measurements directly at the patient connection. The system includes a new REE variable calculated using only CO2, allowing REE measurements to be made at any FiO2 level. With a compact size and ease of mobility, the Ultima CCM offers easy access in otherwise crowded ICU rooms.

Clinical Benefits
  • Assesses outcomes of critically ill patients to provide cost effective health care delivery.
  • Optimizes nutritional needs of thermal injury and trauma patients.
  • Carefully monitors patients with metabolic and eating disorders.
  • Cost-effective management of mechanically ventilated patients reduces the number of vent days and time spent in the ICU.
  • Monitors hemodynamic profiles using the Direct Fick CO in the ICU or Cath Lab.
BREEZESUITE Software

BREEZESUITE comprehensive software package comes with integrated features to enable custom configurations and future expandability. The open architecture database to provide instant data access and robust network connectivity. Helpful coaching features walk the user through simple steps for system calibration and patient testing. BREEZESUITE allows you to design custom reports and configurations, using Report and Script Designer features. Physician Review Station features full trending of all variables for enhanced patient management. BREEZESUITE is compatible with Windows 2000™ and XP Professional™.

 
Changing the Rules for
Nutritional Assessment

 

 

 

Specifications:

 

Innovative Technology

  • Direct Connect preVent Pneumotach provides True Breath-by-Breath measurement under all testing conditions
  • Patented Breath-by-Breath analysis and rapid response O2 and CO2 analyzers.
  • Lightweight (26 gm/<1 oz.) preVent pneumotach exceeds ATS/ERS performance standards

Flow Device

  • Bi-directional Pitot Tube Pneumotach
  • Patent Number: #5,038,773 & #5,119,825
  • Range: ± 18 L/sec
  • Accuracy: ± 3% or 50 ml, whichever is greater (meets or exceeds ATS/ERS clinical performance standards)
  • Resolution: 8.64 ml/sec
  • Deadspace: 39 ml

Flow Device Direct Connect preVent

  • Bi-directional Pitot Tube Pneumotach
  • Patent Number: #5,357,972
  • Accuracy: ± 3% or 10 ml, whichever is greater
  • Resolution: 2.4 ml/sec
  • Deadspace: 14 ml

UL    CE

All specifications are subject to change without notice. Products may vary from those illustrated. Please contact your Medical Graphics representative for the latest information, pricing and product availability.

© 2005-2006 Medical Graphics Corporation. Ultima and preVent are registered trademarks of Medical Graphics Corporation.

Software Options

  • Spirometry
  • Bronchial Provocation
  • Database Query
  • Multi-User Networking

O2 Analysis

  • Patent Number 4,995,256
  • Type: Zirconia
  • Range: 0 - 100%
  • Response: (10-90%) <80 msec
  • Accuracy: ± 0.03%

CO2 Analysis

  • Type: NDIR
  • Range: 0-10%
  • Response: (10-90%)<130msec
  • Accuracy: ± 0.05%

Size (Base)

  • Height: 36 cm (14”)
  • Width: 33 cm (13”)
  • Depth: 36 cm (14”)
  • Weight: 12 kgs (26.5 lbs)

Gas Sample

  • Patent Number 5,042,500
  • Patented gas drying sample circuit
  • Side stream sampling flow rate: 80-100ml./min
  • Warm-up Time: 30 minutes from cold start.

Power Requirements

  • 100-240V/50-60Hz