Arm Adjustments:
- Vertical Extension: 51 cm (20")
- Horizontal Extension: 35.5 cm (14")
- Radius of Extension: 180°
Internal Volume
Size
- Height: 175 cm (69.5")
- Circumference: 287 cm (113")
- Diameter (Door closed) 93 cm (37")
Weight
Gas Requirements
- Compressed air (20 psi)
- 100% Helium (10-28 psi)
(Elite DL & Dx Models)
- DLCO mix (135 psi)
(Elite DL & Dx Models)
Ne: 0.5%
CO: 0.3%
O2: 21%
N2: balance
- 100% Oxygen (135 psi) Elite Dx Model
Models
- Elite D (Spirometry, TGV & Raw)
- Elite DL (Spirometry, TGV, Raw, & DLco)
- Elite Dx (Spirometry, TGV, Raw, DLco,
& N2 washout lung volumes)
Construction
Power Requirements
- 100V/50Hz@2.5A
- 110V/50Hz@2.5A"
- 230V/50Hz@1.75A
- 100V/60Hz@2.5A
- 110V/60Hz@2.5A

|
Flow Device
- Bi-directional differential pressure
- preVent pneumotach
- Patent Number: #5,038,773 & #5,119,825
- Range: ± 18 L/sec
- Accuracy: ± 3% or 50 ml, whichever is greater (meets or exceeds
ATS/ERS clinical performance standards)
- Resolution: 8.64 ml/sec
- Deadspace: 39ml
MedGraphics N2 Analyzer
(Elite DX model)
- Type: Emission spectrophotometry
- Patent Number: #4,898,465
- Range: 0-80%
- Response: <50 msec (10-90% full scale)
MedGraphics Diffusion Analyzer
(Elite DL & Elite Dx Models)
- Type: Gas Chromatography
- Minimum Sample Size: 3 ml

|