Standard Features
Options
- Pulmonary Consult Interpretation Software
- Exercise Consult Interpretation Software
- Bronchial Provocation
- Networking & Information Management
- Database Query
- ExFVL (Exercise Flow Volume Loops)
Flow Device
- Bi-directional Pitot Tube Pneumotach
- Patent Number: #5,038,773 & #5,119,825
- Range: ± 18 L/sec
- Accuracy: ± 3% or 50 ml, whichever is greater
(meets or exceeds
ATS/ERS clinical performance standards)
- Resolution: 8.64 ml/sec
- Deadspace: 39 ml
Gas Sample
- Patent Number 5,042,500
- Patented gas drying sample circuit
- Side stream sampling flow rate: 80-100ml./min
- Warm-up Time: 30 minutes from cold start.
Power Requirements

All specifications
are subject to change without notice. Products may vary
from those illustrated. Please contact your Medical Graphics
representative for the latest information, pricing and product
availability.
©
2005-2007 Medical Graphics Corporation. Ultima and preVent
are registered trademarks of Medical Graphics Corporation.
|
Gas Requirements
- 100% Helium (10 - 28 psi)
- DLco mix (135 psi)
- 100% Oxygen (135 psi)
- Calibration Gas:
- 5% CO2, 12% O2, Bal N2
- 0% CO2, 21% O2, Bal N2
O2 Analysis
- Patent Number 4,995,256
- Type: Zirconia
- Range: 0 - 100%
- Response: (10-90%) <80 msec
- Accuracy: ± 0.03%
Arm Adjustments
- Vertical Extension: 43 cm (17”)
- Horizontal Extension: 35.5 cm (14”)
- Radius: 180º..
Size (Base)
- Height: 36 cm (14”)
- Width: 33 cm (13”)
- Depth: 36 cm (14”)
- Weight: 12 kgs (26.5 lbs)
Size (PF Module)
- Height: 15 cm (6”)
- Width: 33 cm (13”)
- Depth: 30 cm (11.5”)
- Weight: 9.3 kgs (20.5 lbs)
|